MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION ENVEO PRO DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION ENVEO PRO DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number ENVPRO-16
Device Problems Failure to Advance (2524); Patient Device Interaction Problem (4001)
Patient Problem Tachycardia (2095)
Event Date 11/23/2018
Event Type  Injury  
Manufacturer Narrative
Continuation of d10: product id l-envpro-16; product lot/serial number (b)(6); product type: compression loading system; implant date na; explant date na.Product id evolutpro-29; product lot/serial number (b)(6); product type: heart valve; implant date na; explant date na.Select patient information cannot be documented in the regulatory report due to regional privacy regulations.H6.The codes present in section h6 correspond to multiple components/products that comprise this reported event.Analysis: the delivery catheter system (dcs) was received with the valve loaded in the dcs capsule.The dcs was received with the capsule fully closed.The deployment knob retracted and advanced the capsule.The tip-retrieval mechanism was intact.The grey handle was not intact as the front grip separated.The tabs on the front grip were intact with no evidence of damage.The capsule was intact with no evidence of damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the attempted implant of a transcatheter bioprosthetic valve, into extremely tortuous anatomy, the delivery catheter system (dcs) would not advance over the aortic arch.Despite multiple position changes and the use of a buddy wire, the system would not advance and it was withdrawn from the patient.Subsequently, a non-medtronic maneuverable system was used.When the non-medtronic buddy wire was introduced, ventricular tachycardia (vt) was noted which was successfully remedied with an over stimulation by the temporary pacemaker.The vt resolved the same day.The physician indicated the vt was causal to the procedure.The day following the implant of the non-medtronic valve, new left bundle branch block (lbbb) occurred.No treatment was required and the lbbb was reported as resolved with sequelae.The left bundle branch block (lbbb) had not resolved and was still present at the time of the study exit.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVEO PRO DELIVERY SYSTEM
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18647835
MDR Text Key334616308
Report Number2025587-2024-00612
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/03/2020
Device Model NumberENVPRO-16
Device Catalogue NumberENVPRO-16
Device Lot Number0009220926
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Outcome(s) Required Intervention;
Patient Age85 YR
Patient SexFemale
Patient Weight56 KG
-
-