CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number RTLR180343 |
Device Problem
Sparking (2595)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/31/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
It was reported that when the patient plugged their liberty select cycler into the wall outlet it sparked during preparation of their peritoneal dialysis (pd) treatment.There were no recent power outages in the home or in the area reported.There was not an outlet issue.The ok and stop keys did not make a sound when the buttons were pressed.The cycler was switched on but the stop, ok, and up/down keys did not light up.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.Upon follow up, the pdrn confirmed that there were no adverse events or medical intervention required as a result of the reported event.Stated that the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.Confirmed that the patient has received the replacement cycler today.Confirmed that the old cycler is scheduled to be picked up and returned.
|
|
Event Description
|
It was reported that when the patient plugged their liberty select cycler into the wall outlet it sparked during preparation of their peritoneal dialysis (pd) treatment.There were no recent power outages in the home or in the area reported.There was not an outlet issue.The ok and stop keys did not make a sound when the buttons were pressed.The cycler was switched on but the stop, ok, and up/down keys did not light up.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.Upon follow up, the pdrn confirmed that there were no adverse events or medical intervention required as a result of the reported event.Stated that the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.Confirmed that the patient has received the replacement cycler today.Confirmed that the old cycler is scheduled to be picked up and returned.
|
|
Manufacturer Narrative
|
Additional information provided in d9 and h3.Plant investigation: a visual inspection of the returned cycler exterior showed signs of physical damage due to a broken stay safe mount.There were visual indications of dried fluid within the cassette compartment.There were no visual indications of particulates within the cassette area.There were no burrs or sharp edges in cassette area that may have punctured a cassette membrane.Voltage verification check passed.Valve actuation test passed.System air leak test passed.Post-accelerated stress test (ast) 2hr 15mins 8500ml simulated treatment was performed and encountered a grinding motor pump a.The simulated treatment was completed.Mushroom head check passed.An internal visual inspection of the returned cycler was performed.There were visual indications of dried fluid and infestation under the pump assembly on the bottom cover.The cause of the observed dried fluid could not be determined.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.The cycler was refurbished following the evaluation.
|
|
Search Alerts/Recalls
|
|
|