The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturer record evaluation cannot be conducted because the no lot number was provided by the customer.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
It was reported that a patient underwent a ischemic vt (ventricular tachycardia) ablation with a thermocool® smart touch® sf bi-directional navigation catheter and the patient experienced pericardial effusion that required pericardiocentesis.The patient was mapped in tachycardia in the inferolateral area of the left ventricle.A single transseptal puncture was conducted to get to the left side.No steam pops or any other events were noticed during the ablation and it was successful in terminating the tachycardia.While pacing from the st/sf catheter, the body surface signals briefly lost connection which was the only "error" that showed up during the case but no actual errors showed up on carto.During the end of the vt ablation, the patient experienced the pericardial effusion.The ablation procedure was successful, and when the physician was removing catheters they did a final check with the intracardiac echo catheter and discovered the effusion.The effusion was reported to be "unrelated to the ablation procedure." pericardiocentesis was successful and the blood eventually stopped filling.The patient made a full recovery.
|