Catalog Number D134804
Device Problems Signal Artifact/Noise (1036); Patient Device Interaction Problem (4001)
Patient Problem Cardiac Tamponade (2226)
Event Date 01/17/2024
Event Type  Injury  
Manufacturer Narrative
The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturer record evaluation cannot be conducted because the no lot number was provided by the customer.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent a ischemic vt (ventricular tachycardia) ablation with a thermocool® smart touch® sf bi-directional navigation catheter and the patient experienced pericardial effusion that required pericardiocentesis.The patient was mapped in tachycardia in the inferolateral area of the left ventricle.A single transseptal puncture was conducted to get to the left side.No steam pops or any other events were noticed during the ablation and it was successful in terminating the tachycardia.While pacing from the st/sf catheter, the body surface signals briefly lost connection which was the only "error" that showed up during the case but no actual errors showed up on carto.During the end of the vt ablation, the patient experienced the pericardial effusion.The ablation procedure was successful, and when the physician was removing catheters they did a final check with the intracardiac echo catheter and discovered the effusion.The effusion was reported to be "unrelated to the ablation procedure." pericardiocentesis was successful and the blood eventually stopped filling.The patient made a full recovery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18652312
MDR Text Key334682499
Report Number2029046-2024-00399
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberD134804
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN IRRIGATION PUMP; UNKNOWN TRANSSEPTAL NEEDLE; UNKNOWN ULTRASOUND CATHETER; UNK_CARTO 3
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-