Model Number INFKIT2
Device Problem Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned to mmdg for investigation.A dhr review was completed and no non-conformances were found.When the device was returned to mmdg for investigation, the pump alarms operated as expected and were observed visually, however the pump was not able to alarm audibly as a result of the speaker failing.The pump was serviced to correct the issue.
 
Event Description
The initial reporter stated that the pump alarms did not work.They stated that the pump made no audible alarm sound.Mmdg did follow up with the initial reporter who stated that the complaint had occurred during testing and had not had any effect on a patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ENTERAL FEEDING PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin egelhoff
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key18652320
MDR Text Key334884655
Report Number1722139-2024-00096
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberINFKIT2
Device Catalogue NumberINFKIT2
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-