It was reported that the recharger was not taking a charge and wouldn't power up.Visual inspection of the desktop charger (dtc) connector pin and recharger charging port appear to be fine from caller's perspective.The dtc green light is present when plugged into wall.The caller plugged in dtc to recharger and screen came up that didn't show plugged into wall, with a little wiggling and pushing on the connector pin, then icon changed to charging but as soon as caller let go it went way.Sent email to repair for recharger replacement.No symptoms were reported.
|
Continuation of d10: product id 37651 serial# (b)(6) product type recharger product id 37761 lot# serial# (b)(6) product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 37651, serial/lot #: (b)(6), udi#: (b)(4) ; product id: 37761, serial/lot #: (b)(6) h3: analysis of the 37651 recharger (rtm) and 37761 desktop charger (dtc) serial numbers (b)(6) and (b)(6) revealed complaint unverified, no anomalies observed for the rtm and broken connector pins for the dtc.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|