• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. DAFILON BLUE 3/0 (2) 45CM DS19; OTHER SUTURES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. DAFILON BLUE 3/0 (2) 45CM DS19; OTHER SUTURES Back to Search Results
Model Number C0932213
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available a follow-up report will be submitted.
 
Event Description
It was reported an issue with dafilon suture.The client reported that the thread comes off the needle when suturing.Additional information has been requested.
 
Manufacturer Narrative
Summary of investigation: there are no previous complaints of this code-batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical's warehouse.According to the information received, the product was used already expired.We have not received any sample.Without closed samples and/or defective samples a proper analysis cannot be performed.Batch manufacturing record: reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Conclusion root cause analysis: it has not been possible to determine the root cause as no samples have been received.Furthermore, according to the information received the product was used already expired.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done and the case is not confirmed due to lack of evidence.Furthermore, according to the information received the product was used already expired.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Corrective measures: actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 3/0 (2) 45CM DS19
Type of Device
OTHER SUTURES
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key18656871
MDR Text Key334726385
Report Number3003639970-2024-00057
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K151165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/29/2023
Device Model NumberC0932213
Device Catalogue NumberC0932213
Device Lot Number618484
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-