• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Device Difficult to Maintain (3134)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Event Description
It was reported that device failed during preventative maintenance.There was no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
D5.Other operator of device: operator of device is unknown.D9 - date returned to mfg; 01/23/2024.Investigation summary: the affected device returned for investigation was in good physical condition, with no obvious damage present.Water was placed in the water tank, the attached disposable temperature was checked, and the line cord was plugged into the outlet and connected to the electrical analyzer and turned on.However, the customer's reported event could not be replicated; the device passed all electrical and functional testing.As such, the complaint could not be duplicated or confirmed, and a root cause was not found.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.Calibration of the device and preventative maintenance were performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18658734
MDR Text Key334748589
Report Number2183161-2024-00125
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-