• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG EASYPUMP; PUMP, INFUSION, ELASTOMERI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG EASYPUMP; PUMP, INFUSION, ELASTOMERI Back to Search Results
Catalog Number 4540018-02
Device Problem Leak/Splash (1354)
Patient Problem Chemical Exposure (2570)
Event Date 01/10/2024
Event Type  malfunction  
Event Description
As reported by the user facility: brief inquiry description 4540018-02 leaked 5fu on to patient.No injury reported.On (b)(6) 2024 b.Braun melsungen had received an unannounced sample.Technician received clarification from the sales rep that this new sample should be a new complaint.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) sample was submitted to the manufacturer for evaluation.Through visual examination, a crack was observed at the filling port of the sample.The complaint sample was filled with red dye solution to check for leakage.Leakage was observed from the cracked filling port during the filling process.The sample is not within specification; the reported defect is confirmed.A device history record (dhr) review was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.The defect is a result of a failure in the production process.An approved project is in place to further address issues with crack at filling port.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP
Type of Device
PUMP, INFUSION, ELASTOMERI
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
861 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key18660086
MDR Text Key334772676
Report Number9610825-2024-00060
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081905
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4540018-02
Device Lot Number23H08GE561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-