Model Number 471205-17 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
A return material authorization (rma) was issued to the customer requesting to have the intuitive device returned.However, isi has not received the product involved with the alleged issue to perform failure analysis.Additional information is being gathered to determine the contribution of the device to the customer reported issue.A follow-up mdr will be submitted if the product is returned (post failure analysis evaluation).
|
|
Event Description
|
It was reported that after a da vinci-assisted surgical procedure, the fenestrated bipolar forceps instrument was observed to have a burned, cracked plastic at the front of the tip.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the instrument was inspected prior to use.The damage was observed prior to the sterilization.It was difficult to say if the instrument collided with an energy instrument during the procedure.No arcing was observed and no injury to the patient.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and failure analysis investigations replicated/confirmed the customer reported complaint.Failure analysis found charring and localized melting at the yaw pulley.The instrument passed the electrical continuity test.No exposed wire observed.
|
|