Model Number 471205-17
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Manufacturer Narrative
A return material authorization (rma) was issued to the customer requesting to have the intuitive device returned.However, isi has not received the product involved with the alleged issue to perform failure analysis.Additional information is being gathered to determine the contribution of the device to the customer reported issue.A follow-up mdr will be submitted if the product is returned (post failure analysis evaluation).
 
Event Description
It was reported that after a da vinci-assisted surgical procedure, the fenestrated bipolar forceps instrument was observed to have a burned, cracked plastic at the front of the tip.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the instrument was inspected prior to use.The damage was observed prior to the sterilization.It was difficult to say if the instrument collided with an energy instrument during the procedure.No arcing was observed and no injury to the patient.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and failure analysis investigations replicated/confirmed the customer reported complaint.Failure analysis found charring and localized melting at the yaw pulley.The instrument passed the electrical continuity test.No exposed wire observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
FENESTRATED BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18660373
MDR Text Key334775768
Report Number2955842-2024-11158
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471205-17
Device Catalogue Number471205
Device Lot NumberK12230226 0137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-