Model Number 4470 |
Device Problems
Fracture (1260); Use of Device Problem (1670); Defective Device (2588); Contamination /Decontamination Problem (2895)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/18/2024 |
Event Type
Injury
|
Event Description
|
It was reported that this right atrial (ra) lead was explanted due to lead fracture.During the replacement procedure, it was noted that the ra lead insulation was damaged and filled with blood, possibly due to a suture being tied down on the lead instead of the suture sleeve at implant.The ra lead was successfully replaced with a new lead.No additional adverse patient effects were reported.The ra lead is expected to return for analysis.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, visual inspection revealed a cut in the insulation and blood/body fluid in the lumen(s).This type of damage is consistent with induced damage.In this case, we believe the damage occurred during lead implant due to the suture being tied down on the lead and not on a suture sleeve.The reported blood and body fluid in the lead is likely the result of the lead damage observed by the laboratory.
|
|
Event Description
|
It was reported that this right atrial (ra) lead was explanted due to lead fracture.During the replacement procedure, it was noted that the ra lead insulation was damaged and filled with blood, possibly due to a suture being tied down on the lead instead of the suture sleeve at implant.The ra lead was successfully replaced with a new lead.No additional adverse patient effects were reported.The ra lead is expected to return for analysis.
|
|
Search Alerts/Recalls
|