Model Number 44150
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Event Description
It was reported that the filter was not able to be recaptured.The 95% stenosed target lesion was located in the moderately tortuous and mildly calcified left internal carotid artery.A 190 cm filterwire ez was selected for use.During withdrawal, it was noted that the retrieval sheath was not able to fully capture the filter basket.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTERWIRE EZ
Type of Device
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18664143
MDR Text Key334836726
Report Number2124215-2024-02802
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number44150
Device Catalogue Number44150
Device Lot Number0032466572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BALLOON CATHETER: TERUMO.; INFLATION DEVICE: NIPRO.; INTRODUCER SHEATH: DESTINATION.; STENT: CAROTID WALLSTENT.
-
-