• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED ADULT DUAL HEATED BREATHING CIRCUIT; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED ADULT DUAL HEATED BREATHING CIRCUIT; BTT Back to Search Results
Model Number RT200
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2023
Event Type  malfunction  
Event Description
A healthcare facility in china reported that a (b)(6) vented autofeed humidification chamber provided with a rt200 adult dual heated breathing circuit was leaking water during patient use.There were no patient consequences reported.
 
Manufacturer Narrative
(b)(4) section g4: the rt200 adult dual heated breathing circuit is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Section h11: method: the subject mr290v vented autofeed humidification chamber were requested to be provided to fisher & paykel healthcare new zealand, however, neither were provided for evaluation.Our investigation is therefore based on the information provided by the customer and our knowledge of the product.Results: the healthcare facility reported that a (b)(6) vented autofeed humidification chamber provided with a rt200 adult dual heated breathing circuit was leaking water during patient use.Conclusion: without the complaint device or further details on the location of the leakage, we are unable to determine the cause of the reported issue.Every (b)(6) vented autofeed humidification chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject (b)(6) vented autofeed humidification chamber would have met the required specification at the time of production.Our user instructions that accompany the rt200 adult dual heated breathing circuit state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED BREATHING CIRCUIT
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key18665213
MDR Text Key334885599
Report Number9611451-2024-00075
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
SEE H11
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT200
Device Catalogue NumberRT200
Device Lot Number2102555564
Was Device Available for Evaluation? No
Date Manufacturer Received01/09/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age86 YR
Patient SexMale
-
-