PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number AZURION 7 B12
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Event Description
It was reported to philips that the ceiling mount protection cover came off after a collision with the c-arm.No harm to the patient or user was reported.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the issue occurred during a coronary diagnosis procedure and it was completed as planned.The philips field service engineer (fse) inspected the system onsite and confirmed that the shutters were not available and the fluoroscopy was not possible.Review of the log file didn't confirm the reported issue.It was identified that the ceiling mount protection plate and the arm were interfering, and the cover was about to come off.The fse checked and found that the cover of the ceiling mount protector had come off due to the collision.The fse reinstalled the cover that came off to resolve the issue.After the repair, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
derrick massuri
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18666506
MDR Text Key334834781
Report Number3003768277-2024-00745
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K200917
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 B12
Device Catalogue Number722225
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-