• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400 Back to Search Results
Model Number WB91051W
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.The evaluation was unable to duplicate the customers allegation on the multiple error codes.Additionally, the evaluation found the front panel was damaged; the outer socket of the neutral had a major crack; the pictures of the front panel was damaged, and the old software need upgrade.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
It was reported, the endoeye hd ii had multiple errors - error code 433, error code 140, error code 202, error 671 error code 486, mode selection error 140.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR ESG-400
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstraee 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstraee 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18668407
MDR Text Key336061997
Report Number9610773-2024-00376
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/08/2024
1 Device was Involved in the Event
Date FDA Received02/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
-
-