BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/16/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: initial reporter facility name: the device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the twist clamp of the minicap extended life pd transfer set was unable to close; further described as ¿clamp cannot be closed completely." this was observed during use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to h3, h6 and h10.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including leak, clear passage, clamp function testing and torque engagement testing were performed with no issues noted.The reported condition was not verified.The sample met specification.Should additional relevant information become available, a supplemental report will be submitted.
|
|