MEDTRONIC IRELAND ONYX FRONTIER; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number ONYXNG27538UX
Device Problems Device Dislodged or Dislocated (2923); Material Deformation (2976); Positioning Problem (3009)
Patient Problem Insufficient Information (4580)
Event Date 01/24/2024
Event Type  Injury  
Event Description
An attempt was made to use one onyx frontier drug eluting stent to treat a lesion.It was reported that stent dislodgement occurred during removal following a failed delivery.It was detailed that the stent appeared to get stripped off as it was being retrieved back into the guide catheter after it failed to cross the lesion.The dislodged stent was removed with a snare as it dislodged halfway in the vessel and the guide catheter. no further patient injury was reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis summary: the delivery system returned without the stent on the balloon.The dislodged stent returned on a non-mdt snare along with additional non-mdt accessory devices.The dislodged stent was deformed.The balloon folds on the delivery system were partially remained intact and crimp impressions were visible on the exposed balloon surface.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX FRONTIER
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18672360
MDR Text Key334949941
Report Number9612164-2024-00678
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberONYXNG27538UX
Device Catalogue NumberONYXNG27538UX
Device Lot Number0011376456
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-