DEPUY MITEK LLC US VAPR COOLPULSE90 ELECTRODE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Catalog Number 228146 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/31/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.H4: the device manufacture date was unknown.Udi: (b)(4).
|
|
Event Description
|
It was reported by the sales rep in china that during an arthroscopy surgical procedure on (b)(6) 2024 the vapr coolpulse90 electrode device could not suction.The electrode connected to the aspirator and put it into the articular cavity.It was found that there was no suction.Another device was used to complete the surgery, so the aspirator was not the problem.There were no adverse consequences to the patient.No additional information could be provided.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|