• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problems Leak/Splash (1354); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: the device returned to boston scientific consisted of a jetstream xc-2.1 atherectomy catheter.The device was visually examined for any shaft damage.Visual analysis showed shaft damage located 1cm and 90.5cm from the tip.The infusion line showed a burst located 98.5cm from the tip.The device was set up per the instructions for use (ifu) and the device ran as designed.There was a leak noticed at the 98.5cm location.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Device analysis determined the condition of the returned device was consistent with the reported information.The complaint for material deformation, kinks, and a fluid leak were confirmed.
 
Event Description
It was reported that an infusion line leak occurred.A 2.1mm jetstream xc atherectomy catheter was selected for an atherectomy procedure to treat peripheral arterial disease.The superficial femoral artery had been treated successfully; however, when moving distally to the popliteal and posterior tibial trunk, a catheter leak was noticed.There were no patient complications, and an eluvia was placed in the superficial femoral artery.Plain old balloon angioplasty was used to treat the distal portion to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18672960
MDR Text Key336059107
Report Number2124215-2024-06538
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0032560564
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-