• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 6068073
Device Problem Mechanical Problem (1384)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/30/2022
Event Type  Injury  
Manufacturer Narrative
H6: neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a regulatory body/ healthcare professional (hcp) via a medtronic representative regarding a patient implanted with cages in an unknown spinal therapy for an unknown spinal indication.It was reported that the cages collapsed.Levels involved are l5/s1 and l3/4.The date of implant was (b)(6) 2022.The patient experienced pseudoarthrosis on (b)(6) 2022.The date of revision surgery was (b)(6) 2022.The date of last follow up was (b)(6) 2023.No further complications were reported/ anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM
Type of Device
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key18674119
MDR Text Key334948437
Report Number1030489-2024-00095
Device Sequence Number1
Product Code MAX
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K210425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6068073
Device Catalogue Number6068073
Device Lot Number0837725W
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2024
Date Device Manufactured05/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-