Model Number L311 |
Device Problems
Under-Sensing (1661); Material Integrity Problem (2978); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during an implant procedure, this pacemaker showed under sensing at the right atrial (ra) lead channel after the ra lead was placed and screwed into the header.The physician removed the ra lead and connected it up to cables and the pacing system analyzer to check sensing.Sensing and thresholds were appropriate.Re-inserted the same ra lead into the header and the pacemaker still had no sensing.A second device was handed off and appropriate numbers were observed.The ra lead had been in service before this procedure and remains in service at this time.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted dried body fluids consistent with dried blood in the ra port.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during an implant procedure, this pacemaker showed under sensing at the right atrial (ra) lead channel after the ra lead was placed and screwed into the header.The physician removed the ra lead and connected it up to cables and the pacing system analyzer to check sensing.Sensing and thresholds were appropriate.Re-inserted the same ra lead into the header and the pacemaker still had no sensing.A second device was handed off and appropriate numbers were observed.The ra lead had been in service before this procedure and remains in service at this time.The attempted pacemaker has been returned for analysis.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|