MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number MCP00706047#ROTAFLOW ITALIAN EU-PLUG
Device Problems Pumping Stopped (1503); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Manufacturer Narrative
A follow up will submitted when additional information become available.
 
Event Description
It was reported that the rotaflow had a battery issue resulting in a pump stop.The failure occurred during transportation of a patient.No harm to any person has been reported.Complaint id: (b)(4).
 
Event Description
Complaint id: (b)(4).
 
Manufacturer Narrative
It was reported that the rotaflow had a battery issue resulting in a pump stop.The failure occurred during transportation of a patient.No harm has been reported.A getinge field service technician confirmed that the battery pack has been replaced.Thus the failure could be confirmed.The exact root cause remains unknown, but most probably related to battery lifetime.Based on the results the reported failure "battery issue" could be confirmed.The review of the non-conformities has been performed on 2024-02-13 for the period of 2014-06-16 to 2024-01-19.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2014-06-16.In order to avoid reoccurrence of the reported failure, the customer will be informed by the getinge sales and service unit (ssu) to follow the chapter in the instruction for use heart-lung support system rotaflow system| 4.4 | en | v15.Chapter 3.3.4: check battery capacity every 6 months, at the latest.The battery must be replaced by the authorized technical service every 2 years, at the latest.The battery must be replaced sooner if it cannot be fully charged within 8.5 hours or if the system cannot be operated with the fully charged battery.Chapter 5.6.1: before starting the application, check the points listed in "check before every use".Before each use, ensure that the batteries are fully charged.If the battery capacity is low an acoustic signal sounds on the device.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18674710
MDR Text Key334983384
Report Number8010762-2024-00079
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCP00706047#ROTAFLOW ITALIAN EU-PLUG
Device Catalogue Number701043296
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/19/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-