Model Number EVOLUTR-29-C |
Device Problems
Fluid/Blood Leak (1250); Incomplete Coaptation (2507); Patient Device Interaction Problem (4001)
|
Patient Problems
Heart Failure/Congestive Heart Failure (4446); Aortic Valve Insufficiency/ Regurgitation (4450)
|
Event Date 01/27/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that approximately seven years three months following the implant of this transcatheter bioprosthetic valve, the patient experienced symptoms of progressive heart failure.A transesophageal echocardiogram was performed and revealed aortic insufficiency due to flail leaflet of the transcatheter aortic valve.Coronary artery disease was identified via coronary angiography.Subsequently, a surgical aortic valve replacement and coronary artery bypass graft surgery was planned.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information received that the patient¿s symptoms had increased.The specific symptoms were not provided.Severe central a ortic insufficiency was identified.Hospitalization was required and a surgical aortic valve replacement and coronary artery bypass graft (cabg) were performed.
|
|