BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G447
Device Problem High impedance (1291)
Patient Problems Fall (1848); Head Injury (1879)
Event Date 01/30/2024
Event Type  malfunction  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that the patient's spouse called boston scientific latitude customer support stating that the patient fell to the ground due to a shock from their cardiac resynchronization therapy defibrillator (crt-d) and injured their head.A review of device date showed no episode was recorded on the day of the patient's fall (i.E., no shock therapy had been provided that particular day).Daily measurements show some high out-of-range shock impedance measurements from the previous month but none since.No other observations were noted.No adverse patient effects were reported.This product remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18678382
MDR Text Key335562826
Report Number2124215-2024-07607
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/13/2024
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number275589
Was Device Available for Evaluation? No
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-