Manufacturer's investigation conclusion: the reported event of a controller exchange due to the primary controller data cable being shut in a door was not confirmed.The heartmate 3 system controller (serial number (b)(6)) was not returned for analysis.Additionally, there were no log files, photos, or any other supporting documentation that would confirm the reported event.Provided information stated that the patient had a controller exchange because the data cable was shut in a door, resulting in a faulty data cable and the controller will not be returned, the root cause of the reported event was unable to be conclusively determined through this analysis; however, per provided information, the controller cable was shut in a door.Incidental finding revealed fluid ingress in the power cable, via submitted photo.Heartmate 3 instructions for use (rev.C) section 7 entitled ¿alarms and troubleshooting¿ and heartmate 3 patient handbook (rev.D) section 5 entitled ¿alarms and troubleshooting¿ addresses how to properly interpret and troubleshoot all system alarms, including power cable disconnect and low voltage alarms.The heartmate 3 patient handbook (rev.D - section 6 "caring for the equipment) describes how to care for and clean all equipment, including the heartmate 3 system controller and its power cables.The heartmate 3 patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.The device history records were reviewed, and the records revealed the heartmate 3 system controller (serial#: (b)(6)) was manufactured in accordance with manufacturing and qa specifications.No further information was provided.The manufacturer is closing the file on this event.
|