Philips received a complaint on a heartstart mrx device indicating that the device was unable to charge fully.Through a request for additional information, the fse indicates a diagnostic test was ran and found the handle broken and the discharge test was found at a lower value.Insufficient details available to confirm the issue and repair needed.The customer was given information to contact a dealer to purchase a new device.The customer was also informed that service could no longer be completed on the unit as the device had reached end of life (eol).The heartstart mrx device and all associated service/support was discontinued on (b)(6) 2022.Based on the information available there is insufficient information to confirm the reported problem.The customer was made aware of the end-of-life terms and the device remains at the customer site.The investigation concludes that no further action is required at this time.
|