AIZU OLYMPUS CO., LTD. EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-HQ190
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2024
Event Type  malfunction  
Event Description
It was reported, the gastrointestinal videoscope connection for reprocessing is loose.The issue occurred during reprocessing.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found air/water cylinder has no color, suction cylinder has no color, the cover of light guide bundle is damaged, mouthpiece is loose scope connector case unit has a scratch, label on scope connector is damaged.In addition, due to wear of angle wire, bending angle in up direction does not meet the standard value, probe cover has a burn, objective lens has a scratch and universal cord has a wrinkle.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the water supply connector being loose could not be determined.It is likely the event may occur if excessive force is applied to water supply connector when water container¿s metal tip is detached from the connector.Operation manual "preparation and inspection_ inspection of the endoscope" "inspect the control section and the endoscope connector for excessive scratching, deformation, loose parts, or other irregularities".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18682838
MDR Text Key336302994
Report Number9610595-2024-02895
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-HQ190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-