Model Number GIF-HQ190 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/11/2024 |
Event Type
malfunction
|
Event Description
|
It was reported, the gastrointestinal videoscope connection for reprocessing is loose.The issue occurred during reprocessing.There were no reports of patient harm.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation and the evaluation found air/water cylinder has no color, suction cylinder has no color, the cover of light guide bundle is damaged, mouthpiece is loose scope connector case unit has a scratch, label on scope connector is damaged.In addition, due to wear of angle wire, bending angle in up direction does not meet the standard value, probe cover has a burn, objective lens has a scratch and universal cord has a wrinkle.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the water supply connector being loose could not be determined.It is likely the event may occur if excessive force is applied to water supply connector when water container¿s metal tip is detached from the connector.Operation manual "preparation and inspection_ inspection of the endoscope" "inspect the control section and the endoscope connector for excessive scratching, deformation, loose parts, or other irregularities".Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|