H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one maxcore disposable core biopsy instrument was returned for evaluation.During visual testing, device appeared to have residue throughout the outer housing and received with both slides in their initial position.A slight bend was observed to the needle.No functional testing was performed due to the nature of the complaint.Two electronic photo was provided by the customer and reviewed.Both the photos show the maxcore device needle in which the sample notch is exposed, and it was noted to be bent.Based on the photo review, the reported needle bent can be confirmed and difficult to remove is unconfirmed.Therefore, based on the provided photos and sample testing the investigation for the reported needle bent can be confirmed as a slight bend was observed to the needle and the difficult to remove issue remains inconclusive as functional testing was not performed.A definitive root cause for the alleged needle bent and difficult to remove issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 06/2024) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
It was reported that during an ultrasound guided lymph biopsy through normal density tissue, the needle of the device was allegedly bent.It was further reported that the device was allegedly difficult to remove.Reportedly, no coaxial was used and the procedure was completed with another device.There was no reported patient injury.
|