|
Model Number IPN048826 |
Device Problem
Complete Blockage (1094)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that "the liquid does not come out of the syringe and the device cannot used.Moreover, the nasal adaptor broke".The patient was reported as fine post the procedure.Associated complaints 3003898360-2024-00211.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint of "occlusion - blockage/part stuck" was confirmed based upon the sample received.A device history record review was performed on the device with no evidence to suggest a manufacturing related root cause.Visual examination of the returned device revealed that the nasal plug and distal tip were detached from the lumen tube.Although the device would be unable to function properly with the distal tip detached, an attempt to fire the device was made to see if anything was able to fire.The returned syringe was filled with water and then attached to the lumen tube to test for an occlusion of the returned device.Upon testing, the device was not able to fire any water.It appeared that there was an occlusion in the device.The manufacturing site inspected the device and found no issues with the positioning of the insert (component that helps with atomization).It could not be determined what prevented the device from atomizing properly at this time.The qa inspector performs a random atomization test in the press process during the manufacturing of this device to help identify this issue, if it occurs, at the time of manufacturing.Teleflex will continue to monitor and trend on this issue.
|
|
Event Description
|
It was reported that "the liquid does not come out of the syringe and the device cannot used.Moreover, the nasal adaptor broke".The patient was reported as fine post the procedure.Associated complaints 3003898360-2024-00211 and 3003898360-2024-00212.
|
|
Search Alerts/Recalls
|
|
|