Model Number L311
Device Problems Failure to Capture (1081); Pacing Problem (1439); Failure to Read Input Signal (1581); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2024
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this pacemaker was under sensing the right atrial lead channel.Sometimes the pacemaker atrial paces right on top of a conducted ventricular sensed deflections.There was low amplitude observed.The right ventricular automatic capture was in suspension.This was possibly due to ventricular sensing falling in blanking, so the pacemaker ventricular paces and there was functional loss of capture due to the conducted ventricular sense in blanking period.The patient was going to look for a cardiologist to determine the process to follow.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI DR
Type of Device
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18683407
MDR Text Key335091616
Report Number2124215-2024-07809
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number700337
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-