The manufacturer received information stating the trilogy 100 ventilator did not meet acceptance criteria of evaluation process for the following conditions: o2 port broken.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, the device powered on and operated as it should.There were no significant event (s) in the error log.It was confirmed that the o2 port was broken, and screw hubs cracked.The trilogy rear case kit, enclosure seam gasket and trilogy plastic o2 connector were replaced to address the issue.The inlet air path assembly and removable air path foam were replaced per remediation. the device was sent for further evaluation.The device failed pressure sensor calibration.The sensor board pca assembly was replaced to address the issue.In addition, talc was found in the transition tube and flow sensor assembly.All components were replaced.
|