An event of a device deforming during implantation was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined, however is consistent with use of larger sheath, as use of a larger sheath may influence deformations of the device.There is no indication of a product quality issue with regards to manufacture, design, or labeling.Please note that per the instructions for use, the minimum recommended size delivery system for use with a 25-18mm amplatzer talisman pfo occluder is an 8f.
|