Model Number 816571
Device Problems Mechanical Jam (2983); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) checked for over occlusion and cleaned the optical sensor but the reported issue remained.He replaced the pump pod and the issue was resolved.The unit operated to the manufacturer's specifications.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump displayed a 'pump jam' message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed that when the pod subassembly was installed into a lab use only (luo) roller pump, pump jam errors occurred as soon as the pump was started, even without tubing or occlusion.He observed the resistor at r31 of the application board was loose.The solder termination had been broken and the unattached end of the resistor was floating above its attachment point.This would impact the output of the circuit which connects to the roller pump gut tachometer signal causing persistent pump jam messages.Once the resistor was repositioned to complete the circuit the pump jam errors stopped occurring.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key18687806
MDR Text Key335207325
Report Number1828100-2024-00051
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816571
Device Catalogue Number816571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/05/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-