A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and the fse did a functional check with patient simulator and intra-aortic balloon, after more than 15 minutes of operation, the cardiosave did not generate any alarm.Users told the fse that the intra-aortic balloon used is from the arrow teleflex brand (ultraflex iab / 7.5 fr / 2.5mm /40 cc).The service did not keep the arrow intra-aortic balloon therefore impossible to request an expertise.Functional check with patient simulator and the iapb consumable under the same conditions as the user settings.Functional control with the different simulators without having any alarm.Checking the ecg cable.Leakage test of the safety disk (pim), drive and fill manifold test of the compressor.Analysis of log errors.Verification of the correct functioning of the leak alarms by simulating them.New functional check for 30 minutes, there was no alarm.After the various analyzes and the various tests, the cpbia cardiosave console does not presents no problem.Equipment tested according to the manufacturer and operational protocol.
|