Model Number 0998-00-0800-55
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Event Description
It was reported that during use on a patient, a gas leak occurred during the night of (b)(6) 2024 to (b)(6) 2024 on the cardiosave intra-aortic balloon pump (iabp).The problem is persistent.There was no patient harm or injury reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and the fse did a functional check with patient simulator and intra-aortic balloon, after more than 15 minutes of operation, the cardiosave did not generate any alarm.Users told the fse that the intra-aortic balloon used is from the arrow teleflex brand (ultraflex iab / 7.5 fr / 2.5mm /40 cc).The service did not keep the arrow intra-aortic balloon therefore impossible to request an expertise.Functional check with patient simulator and the iapb consumable under the same conditions as the user settings.Functional control with the different simulators without having any alarm.Checking the ecg cable.Leakage test of the safety disk (pim), drive and fill manifold test of the compressor.Analysis of log errors.Verification of the correct functioning of the leak alarms by simulating them.New functional check for 30 minutes, there was no alarm.After the various analyzes and the various tests, the cpbia cardiosave console does not presents no problem.Equipment tested according to the manufacturer and operational protocol.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18687830
MDR Text Key335248787
Report Number2249723-2024-00571
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ARROW TELEFLEX BRAND ULTRAFLEX IAB
-
-