During an atrioventricular nodal reentrant tachycardia case, there was a delay due to the ampere rf generator.The tacticath df ablation catheter was positioned at the location of the slow pathway and as soon as the power on button was pressed for ablation, an alarm sounded saying impedance out of range.This was despite the impedance displaying a value of only 130 ohms.Connections to the ampere rf generator were confirmed and the auto flow settings were reset as they had reverted to the default.Ablation was attempted again with the same outcome.A replacement ground was used to resolve the issue but again even though the impedance was below the 150 ohm setting, no ablation could be delivered.A flexability se catheter was then connected to the ampere rf generator, using a different ablation extension cable.Again, the impedance was registering below the impedance cutoff of 150 ohms, but ablation could not be delivered because of the impedance issue.It was suggested to reboot the ampere rf generator which had been done previously but it was also asked which ablation ground was connected to the patient and ampere rf generator.It was a non-abbott device (conmed 410-2000) which was said to not be compatible with the ampere rf generator.However, this ground had been used for the last 18 months.A competitive ablation catheter with the stockert rf generator was used in order to complete the case successfully.
|