TENEX HEALTH, INC. TX SYSTEM - TX2 MICROTIP; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 554-2003-001
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2024
Event Type  malfunction  
Event Description
During a procedure with the tenex system, a portion of the microtip needle separated from the rest of the handpiece.There were no patient complications.
 
Manufacturer Narrative
The device was returned and evaluated.It was confirmed that part of the microtip needle had separated from the rest of the handpiece.The fracture site did not immediately indicate a specific cause for the failure.Testing and disassembly of the device did not reveal any further anomalies or defects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TX SYSTEM - TX2 MICROTIP
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630 8123
Manufacturer (Section G)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630
Manufacturer Contact
david vancelette
26902 vista terrace
lake forest, CA 92630
9492388220
MDR Report Key18691266
MDR Text Key335534920
Report Number1000135560-2024-00007
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153299
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number554-2003-001
Device Catalogue Number554-2003-001
Device Lot Number2842301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
Patient Weight91 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-