Model Number 1054260 |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/22/2024 |
Event Type
malfunction
|
Event Description
|
A ventilator was returned to the manufacturer for foam remediation.There was no harm or injury reported.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, the device failed test steps during testing for the hw power fail test.The active exhalation control module was replaced to resolve this issue on the device.The internal/detachable battery was replaced on the device.Additional findings not related to the malfunction/issue include the screw beneath porting block was not seated, damage to the bottom on the front enclosure.The customer does not want any repairs done to the unit and it will be returned unrepaired.
|
|