It was reported this was an arteriotomy closure of the left common femoral artery using the pre-close technique via a 6f sheath hole prior to a correction of iliac pseudo aneurysm with endoprosthesis extension interventional procedure.Reportedly, a click was not heard when deploying the needle plunger of the proglide device and "it did not stay in position".The plunger was pressed to its limit and did not make the usual click.When relieving pressure on the plunger, it spontaneously returned almost 1 cm.The plunger was pushed three more times and not once did it click and it continued to return spontaneously the same distance.When the plunger was removed, a cuff miss [suture retrieval issue] was noticed.The sutures of two new proglide devices were successfully pre-placed.The sheath was upsized to a 14f sheath and the interventional procedure was completed.Hemostasis was achieved with the pre-placed proglide sutures.There was a 15-minute delay to replace the device; however, there was no reported adverse patient sequela.No additional information was provided.
|
The device was not returned for analysis.A review of the manufacturing records identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.Femoral imaging was not performed.It should be noted that the electronic perclose proglide instructions for use (ifu), states: perform a femoral angiogram to verify the location of the puncture site.The ifu deviation did not contribute to the reported difficulties as calcification was not found and successful access was achieved.The cause of the reported difficulties cannot be determined.The subsequent treatments are due to the circumstances of the procedure.In this case, it is possible that needle to cuff miss occurred jamming the needle likely against the foot thus prohibiting the full compression of the plunger.In this condition, the needles of the device would be in a bend inducing the spring back (unintended motion) observed.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|