Model Number 471172-17 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/16/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during a da vinci-assisted surgical procedure, the maryland bipolar forceps (mbf) instrument pulley cover had a thermal damage.The customer used a backup mbf instrument to continue with the planned procedure.No fragment fell inside the patient.No known impact or patient consequence was reported.Intuitive surgical, inc.(isi) followed up with the site and obtained the following additional information: the mbf instrument was inspected prior to use with no issue.The customer observed charred tissue on the instrument.There was no report of the tissue being burned unexpectedly.The instrument did not collide with any energy instrument during the procedure.There was no arcing observed during the procedure.There was no patient injury.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) has not received the maryland bipolar forceps instrument for evaluation.Therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) received the maryland bipolar forceps instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the reported complaint.The instrument was found to have charring and localized melting at the grip base between the grips.The instrument passed the electrical continuity test.The instrument was placed on an in-house system, and no arcing was observed.
|
|
Search Alerts/Recalls
|