Related manufacturing reference number: 2017865-2024-22425.It was reported that the patient presented for an ablation procedure and device changeout procedure.Prior to the procedure, it was noted that the right ventricular lead exhibited noise oversensing.The physician explanted and replaced the lead.During the procedure, it was noted that the new right ventricular lead was unable to sit properly in the header of the new implantable cardioverter defibrillator (icd).The icd was removed and replaced and the lead was successfully attached.The patient was stable.
|
The reported event of failure to connect and insertion anomaly could not be confirmed.Final analysis found the device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
|