INTUITIVE SURGICAL, INC ENDOWRIST; MARYLAND BIPOLAR FORCEPS Back to Search Results
Model Number 471172-17
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Event Description
It was reported that after a da vinci-assisted surgical procedure, the maryland bipolar forceps instrument had a damaged black insulation about three centimeters from the tip.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the maryland bipolar forceps instrument damage was observed prior to sterilization.There were no exposed wires and no broken cables.There were signs of thermal damage, but no arcing was observed.The customer believed that the instrument collided with another instrument or tool.There were no issues removing the instrument from the cannula, and the procedure was completed with the same instrument.
 
Manufacturer Narrative
An investigation is in progress to determine the cause if the reported event.An return material authorization (rma) was issued to evaluate the intuitive surgical, inc.(isi) device.Additional information is being gathered to determine the contribution of the device to the customer reported issue.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The maryland bipolar forceps instrument was analyzed and found to have damage on the conductor wire¿s insulation at the yaw pulley.There was material missing and the conductor wires were exposed.The conductor wire insulation was damaged, but the wire was not torn or fully broken.The instrument passed an electrical continuity test.Components adjacent to the damaged wire insulation do not show damage.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
MARYLAND BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18692528
MDR Text Key335184260
Report Number2955842-2024-11214
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471172-17
Device Catalogue Number471172
Device Lot NumberK10221128 0237
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-