Model Number 471172-17 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that after a da vinci-assisted surgical procedure, the maryland bipolar forceps instrument had a damaged black insulation about three centimeters from the tip.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the maryland bipolar forceps instrument damage was observed prior to sterilization.There were no exposed wires and no broken cables.There were signs of thermal damage, but no arcing was observed.The customer believed that the instrument collided with another instrument or tool.There were no issues removing the instrument from the cannula, and the procedure was completed with the same instrument.
|
|
Manufacturer Narrative
|
An investigation is in progress to determine the cause if the reported event.An return material authorization (rma) was issued to evaluate the intuitive surgical, inc.(isi) device.Additional information is being gathered to determine the contribution of the device to the customer reported issue.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The maryland bipolar forceps instrument was analyzed and found to have damage on the conductor wire¿s insulation at the yaw pulley.There was material missing and the conductor wires were exposed.The conductor wire insulation was damaged, but the wire was not torn or fully broken.The instrument passed an electrical continuity test.Components adjacent to the damaged wire insulation do not show damage.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|