MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH S5 ERC TUBING CLAMP / 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH S5 ERC TUBING CLAMP / 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-05-65
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in usa.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland has received a report that, during a procedure, the erc shuts off.The erc is not registering with the pump.There was no report of any patient injury.
 
Manufacturer Narrative
As per information provided from fse on site for pm, the complained unit was re-tested by hospital personnel and the issue was reproduced.The erc was removed from service and scrapped by customer, therefore further testing of the device was not possible.A device service history review has been performed and identified that the unit was manufactured in 2015 and no other similar event has been reported, neither concerning trend has been identified.The specific root cause for the mentioned occurrence is unable to be identified but, considering the results of a similar complaints review and the type of failure, it cannot be ruled out that it is the failure of one of the internal boards, most likely the zka board where the processor, the connector for can connection cable to s5 and the connector to the zkb motor board are present.Failure of electronic boards can be related to multiple and not deterministic factors such as exposure to heat, dust and moisture, accidental impact (drops and falls), and power overloads/surges but also to variability of micro subcomponents.However, there is no concerning trend for this kind of failure.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ERC TUBING CLAMP / 620MM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18694623
MDR Text Key335245569
Report Number9611109-2024-00077
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-05-65
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-