Catalog Number 03P89-24
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2024
Event Type  malfunction  
Event Description
On 02-feb-2023, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded suspected discrepant inr results a 63 year old female patient with deep vein thrombosis.There was no additional nformation available at the time of this report.Return product is available for investigation.Method date collected tested results (sec) sample i-stat (b)(6) 2024 11:46 11:50 *>8 5 acl top (b)(6) 2024 12:50 13:50 4.1 6 at this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.Pt/inr - intended use the i-stat pt, a prothrombin time test, is useful for monitoring patients receiving oral anticoagulation therapy such as coumadin ® or warfarin.
 
Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on 09-apr-2024.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained and returned cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.An (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT PT/INR CARTRIDGE
Type of Device
PT/INR CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136123525
MDR Report Key18696283
MDR Text Key335259509
Report Number2245578-2024-00021
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020355
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03P89-24
Device Lot NumberF23323
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-