Model Number 0600-0783
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Event Description
A customer reported that during a patient procedure, using a glidescope core smart cable, the picture would either not display or it would freeze.The procedure was completed using a backup cable which was made available in an unspecified amount of time.No delay in the procedure or harm to the patient was reported.
 
Manufacturer Narrative
The customer declined the option to have their glidescope core smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Review of complaint history for the reported smart cable serial number "(b)(6)" did not identify any previous complaints reported to verathon.Trending analysis for the glidescope core smart cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key18697283
MDR Text Key336421286
Report Number9615393-2024-00022
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0600-0783
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/11/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-