BOSTON SCIENTIFIC CORPORATION ACUITY X4 STRAIGHT; IMPLANTABLE LEAD Back to Search Results
Model Number 4671
Device Problems High impedance (1291); Positioning Problem (3009); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2024
Event Type  malfunction  
Event Description
It was reported that this left ventricular (lv) lead was an attempted implant due to high impedance measurements and high capture thresholds.The physician opted to replace the lead and a new one was successfully implanted.No adverse patient effects were reported.This lead has been received for analysis.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
 
Event Description
It was reported that this left ventricular (lv) lead was an attempted implant due to high impedance measurements and high capture thresholds.The physician opted to replace the lead and a new one was successfully implanted.No adverse patient effects were reported.This lead has been received for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY X4 STRAIGHT
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18699145
MDR Text Key336396029
Report Number2124215-2024-08446
Device Sequence Number1
Product Code LWP
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S398
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number4671
Device Catalogue Number4671
Device Lot Number873113
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
-
-