PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING Back to Search Results
Model Number 865350
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2024
Event Type  malfunction  
Event Description
The customer biomedical engineer reported there was a speaker malfunction, and confirmed there was no sound at all.The device was not in clinical use at the time the issue was discovered.There was no adverse event or patient harm reported.
 
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
The device was sent to philips authorized repair facility (rft) for bench for evaluation.Results of functional testing indicate that speaker produced audible sound.Based on the information available and the testing conducted, the cause of the reported problem was not replicated.The reported problem was not confirmed.Although the speaker was confirmed to be functioning per specification during testing it was indicated that there was sound at the time of the event, the speaker has been replaced per current process.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18699186
MDR Text Key335324298
Report Number1218950-2024-00095
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-