Model Number L211
Device Problems Use of Device Problem (1670); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2024
Event Type  malfunction  
Manufacturer Narrative
The product is expected to be returned for analysis.This report will be updated upon return and completion of analysis.
 
Event Description
It was reported that during a lead revision procedure, a non-boston scientific torque wrench was used to loosen the setscrew for this pacemaker.The wrench became stuck and when trying to pull the wrench out, the setscrew was pulled out of the device header.The setscrew was eventually able to be freed from the wrench, but due to the damage, a non-boston scientific pacemaker was implanted following the lead repositioning.No adverse patient effects were reported.The pacemaker was expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPONENT MRI DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18699277
MDR Text Key336396287
Report Number2124215-2024-08453
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberL211
Device Catalogue NumberL211
Device Lot Number707254
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2023
Is the Device Single Use? Yes
Type of Device Usage Initial