SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problems
Display or Visual Feedback Problem (1184); Low Readings (2460)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3.Date of event and d5.Operator of device: unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.
|
|
Event Description
|
It was reported that the product heated up, but the temperature display on the main unit showed 10 to 11 degrees.There was no patient involvement.
|
|
Manufacturer Narrative
|
One device was received for investigation.The device was functionally tested.Although it heated up, it was confirmed that the temperature display on the main body lcd did not display the correct value and the led did not light up.Failure of the main pc board inside the main unit is the root cause of the problem.The main pcboard and ac power switch need replacement.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
|