• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems Display or Visual Feedback Problem (1184); Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3.Date of event and d5.Operator of device: unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.
 
Event Description
It was reported that the product heated up, but the temperature display on the main unit showed 10 to 11 degrees.There was no patient involvement.
 
Manufacturer Narrative
One device was received for investigation.The device was functionally tested.Although it heated up, it was confirmed that the temperature display on the main body lcd did not display the correct value and the led did not light up.Failure of the main pc board inside the main unit is the root cause of the problem.The main pcboard and ac power switch need replacement.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18699873
MDR Text Key335390723
Report Number3012307300-2024-00776
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/17/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received02/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-