Model Number PCF-H190DL |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/09/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device has been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.The customer provided the cleaning sterilization and disinfection (cds) processes performed at the user facility.The customer did confirm that there were no deviations or deficiencies concerning reprocessing of the scope.Additionally, the customer confirmed that there were no suspected patient infections due to the facility findings.The customer did not provide the specific steps taken during the cleaning sterilization and disinfection (cds) process.The device evaluation found the distal end insulation test failed; the angulation was out of specification; the image flared; the light guide lenses were worn and chipped; and the adhesive on bending section cover was detached, chipped, and cracked.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
It was reported that, during reprocessing, the colonovideoscope tested positive for <10 colony forming units (cfus) of pseudomonas on january 9, 2024 and january 15, 2024.The device was sampled again on january 24, 2024, and tested negative.It was not specified which channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer completed cleaning disinfection and sterilization (cds) checklist and the lms final investigation.The lm reviewed the customer provided the cds processes where information to judge deviation from the instructions for use (ifu) was not identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that reprocessing was insufficiently conducted.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|