Catalog Number 05P58-25
Device Problem False Positive Result (1227)
Patient Problems Abdominal Pain (1685); Nausea (1970); Vomiting (2144); Dysuria (2684); Decreased Appetite (4569)
Event Date 03/23/2024
Event Type  malfunction  
Event Description
On (b)(6) 2024, abbott point of care (apoc) was contacted by a customer regarding i-stat b-hcg cartridges that yielded a positive result on a 18 year old female patient with abdominal pain, decreased appetite, nausea, vomiting, dysuria.There was no additional patient information at the time of this report.Return product is not available for investigation.Method date collected tested result i-stat on (b)(6) 2021 09:56am 10:08am 25.1 iu/ml roche on (b)(6) 2021 (b)(6) <5.0 iu/ml there are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.I-stat positive cut-off values: b-hcg < 5.0 iu/l negative 5.0 < ss-hcg < 25.0 iu/l indeterminate b-hcg >25.0 iu/l positive intended use: the i-stat® total beta-human chorionic gonadotropin (b-hcg) test is an in vitro diagnostic test for the quantitative and qualitative determination of b-hcg in venous whole blood or plasma samples using the i-stat 1 analyzer systems.The test is intended to be used as an aid in the early detection of pregnancy and is for prescription use only.
 
Manufacturer Narrative
Apoc incident (b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.
 
Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 24-apr-2024.A review of the device history record confirmed the cartridge lot met finished goods release criteria.Retained cartridges could not be tested as the cartridge lot expired on 14-may-2021, prior to the communication of the discrepant result on (b)(6) 2024.No deficiency has been identified for total b-hcg cartridge lot f20333.
 
Event Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT B-HCG CARTRIDGE
Type of Device
B-HCG CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
185 corkstown road
princeton, NJ K2H 8-V4
CA   K2H 8V4
6136123525
MDR Report Key18702696
MDR Text Key335355897
Report Number2245578-2024-00043
Device Sequence Number1
Product Code DHA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/14/2021
Device Catalogue Number05P58-25
Device Lot NumberF20333
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age21 YR
Patient SexFemale
-
-