• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID AND BLOOD WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID AND BLOOD WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problems No Audible Alarm (1019); Alarm Not Visible (1022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.D5.Other operator of device: operator of device is unknown.D9.Date returned to mfg: unknown.No information has been provided to date.H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation summary: the affected device was received for investigation, with a marked-up enclosure and faded pole clamp.After visual inspection, the water tank was filled with water, the temp check disposable was attached, the line cord was plugged in, and the power switch was turned on.The over temp alarm had no led light or audible alarm, confirming the customer's indicated failure.The root cause was determined to be a faulty pcb.As a result, the pcb and reservoir gasket were replaced, and the enclosure was cleaned of marks and scratches.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Event Description
It was reported that while testing, there was no visible alarm.There was no patient involvement and no patient adverse/harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID AND BLOOD WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18702820
MDR Text Key336426435
Report Number2183161-2024-00131
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-390
Device Catalogue NumberHL-390
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse